Menu
Contact
Home / Equipment / Promotions, Special Prices & Clearances / Belmont Touch Phot-X II Intraoral X-Ray – Stock Clearance

Belmont Touch Phot-X II Intraoral X-Ray – Stock Clearance

SKU: EE7756-SC

Belmont Touch Phot-X II Intraoral X-Ray – Stock Clearance

SKU: EE7756-SC

  • Overview
  • Description
  • Specifications
  • Downloads
Overview

The Belmont Touch Phot-X II is an intraoral X-ray system designed for clear imaging, flexible positioning and straightforward touchscreen control.

Its wall-mounted glass touchscreen provides access to patient presets, manual exposure settings, multilingual operation, USB traceability and error reporting. The zero-drift scissor arm helps keep the X-ray head stable during positioning, while selectable 60 kV or 70 kV output supports different imaging requirements.

The system is compatible with digital sensors and film, with multiple arm lengths and mounting options available to suit different surgery layouts.

Available as a stock clearance unit while stocks last. Price on application.

Description

Touchscreen Control

The Belmont Touch Phot-X II features a wall-mounted touchscreen panel with a smooth glass surface and icon-based interface.

The control panel includes sleep mode, multilingual operation, USB connectivity for traceability and instant error reporting.

It can be positioned outside the surgery to support safer operation and easier separation between the operator and the X-ray unit.

Stable Scissor-Arm Positioning

The zero-drift scissor arm is designed to keep the X-ray head securely in position once placed.

This helps support consistent image capture while reducing the need for repeated adjustments during treatment.

Three arm lengths are available to suit different room sizes and installation requirements.

Flexible Mounting Options

The Phot-X II can be configured for wall, ceiling or mobile installation.

A unit with stool configuration is also available, giving practices more flexibility when planning the X-ray position around existing surgery equipment and available space.

Patient-Specific Exposure Settings

Preset exposure options are available for adult, adolescent and child patients.

Manual override allows exposure settings to be adjusted where more precise control is required.

The system is compatible with both digital sensors and film imaging workflows.

Selectable Imaging Output

The Phot-X II provides selectable tube potentials of 60 kV or 70 kV and selectable tube currents of 4 mA or 7 mA.

Exposure times range from 0.01 to 3.2 seconds, allowing the operator to adjust settings across different patient and imaging requirements.

A 0.7 mm focal spot supports clear diagnostic imaging.

X-Ray Tube And Generator

The unit uses a Toshiba D-0711 stationary anode X-ray tube with a tungsten target and 16-degree target angle.

Anode heat content is rated at 7 kJ or 10 kHU, while the tube assembly heat content is 120 kJ or 170 kHU.

The high-voltage generator provides an output of 0.49 kW at 70 kV and 7 mA.

Radiation Filtration And Field Control

The unit includes 1.7 mm aluminium equivalent inherent filtration and 0.3 mm added aluminium filtration.

Minimum filtration is 2.0 mm aluminium equivalent at 70 kV.

Radiation field size is 60 mm diameter with the circular cone or 36 x 47 mm with the rectangular option.

Cone Options

The standard regular cone provides an SSD of 203 mm.

Optional configurations include a 305 mm long cone and a rectangular cone with an SSD of 203 mm.

These options allow the system to be configured around different clinical and positioning requirements.

Built For Regular Clinical Use

The Phot-X II is designed for regular use in busy dental environments.

Its smooth wipeable surfaces support routine cleaning while the recessed focal spot design contributes to a compact and practical X-ray head.

The unit has a duty cycle of 1:50 to manage operating temperature and protect the equipment during repeated use.

Stock Clearance Availability

This Belmont Touch Phot-X II Intraoral X-Ray is available as a stock clearance product.

Stock clearance availability is limited and subject to remaining stock. Contact us for current availability, condition, specification and pricing.

Specification

Specifications:

  • X-ray tube: Toshiba D-0711 stationary anode
  • Focal spot: 0.7 mm
  • Target material: Tungsten
  • Target angle: 16 degrees
  • Anode heat content: 7 kJ / 10 kHU
  • Tube assembly heat content: 120 kJ / 170 kHU
  • Tube potential: 60 kV / 70 kV selectable
  • Tube current: 4 mA / 7 mA selectable
  • Exposure time: 0.01 to 3.2 seconds
  • HV generator output: 0.49 kW at 70 kV, 7 mA
  • Inherent filtration: 1.7 mm Al equivalent
  • Added filtration: 0.3 mm Al
  • Minimum filtration: 2.0 mm Al equivalent at 70 kV
  • Circular radiation field: 60 mm diameter
  • Rectangular radiation field: 36 x 47 mm
  • Cone base distance: 94 mm
  • Regular cone SSD: 203 mm
  • Long cone SSD: 305 mm
  • Rectangular cone SSD: 203 mm
  • Duty cycle: 1:50
  • Horizontal head movement: 0 to 600 degrees
  • Vertical head movement: 0 to 300 degrees
  • Control panel: Wall-mounted touchscreen
  • Patient presets: Adult, adolescent and child
  • Manual exposure override: Included
  • USB traceability: Included
  • Multilingual operation: Included
  • Error reporting: Included
  • Sleep mode: Included
  • Imaging compatibility: Digital sensors and film
  • Arm type: Zero-drift scissor arm
  • Arm length: 300 mm
    • Reach with 300 mm arm: 1,422 mm
  • Arm length: 800 mm
    • Reach with 800 mm arm: 1,922 mm
  • Arm length: 1,000 mm
    • Reach with 1,000 mm arm: 2,122 mm
  • Mounting options: Wall, ceiling, mobile and unit with stool
  • Supply voltage: 230 Vac
  • Supply range: 207 to 253 Vac
  • Line power: 1.5 kVA
  • Line current at 70 kV, 7 mA: 6.6 Aac
  • Maximum line current: 7.3 Aac
  • Internal resistance: Maximum 1.02 ohms
  • Power frequency: 50 Hz single phase
  • Operating temperature: 10 to 40°C
  • Operating humidity: 30 to 75%
  • Operating pressure: 700 to 1,060 hPa
  • Storage temperature: -20 to 70°C
  • Storage humidity: 10 to 90%
  • Storage pressure: 500 to 1,060 hPa
  • Availability: Stock clearance